:FUNCtion:RANGe

(see page 530)
Command Syntax :FUNCtion:RANGe <range>
<range> ::= the full-scale vertical axis value in NR3 format.
The :FUNCtion:RANGe command defines the full- scale vertical axis for the
selected function.
Query Syntax :FUNCtion:RANGe?
The :FUNCtion:RANGe? query returns the current full- scale range value for
the selected function.
Return Format <range><NL>
<range> ::= the full-scale vertical axis value in NR3 format.
The range for ADD, SUBT, MULT is 8E- 6 to 800E+3.
The range for the INTegrate function is 8E- 9 to 400E+3 (depends on sweep
speed).
The range for the DIFFerentiate function is 80E-3 to 8.0E12 (depends on
sweep speed).
The range for the FFT (Fast Fourier Transform) function is 8 to 800 dBV.
See Also "Introduction to :FUNCtion Commands" on page 182