Obsolete and Discontinued Commands 5

:CHANnel2:SKEW

Command Syntax

N O T E

(see page 530)

:CHANnel2:SKEW <skew value>

<skew value> ::= skew time in NR3 format

<skew value> ::= -100 ns to +100 ns

The :CHANnel2:SKEW command sets the skew between channels 1 and 2. The maximum skew is +/- 100 ns. You can use the oscilloscope's analog probe skew control to remove cable delay errors between channel 1 and channel 2.

The :CHANnel2:SKEW command is an obsolete command provided for compatibility to previous oscilloscopes. Use the :CHANnel<n>:PROBe:SKEW command (see page 154) instead.

N O T E

This command is only valid for the two channel oscilloscope models.

 

Query Syntax :CHANnel2:SKEW?

The :CHANnel2:SKEW? query returns the current probe skew setting for the selected channel.

Return Format <skew value><NL>

<skew value> ::= skew value in NR3 format

See Also "Introduction to :CHANnel<n> Commands" on page 143

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference