3Commands by Subsystem

Table 31 Service Request Enable Register (SRE) (continued)

Bit

Name

Description

When Set (1 = High = True), Enables:

 

 

 

 

5

ESB

Event Status Bit

Interrupts when enabled conditions in the

 

 

 

Standard Event Status Register (ESR) occur.

 

 

 

 

4

MAV

Message Available

Interrupts when messages are in the Output

 

 

 

Queue.

 

 

 

 

3

---

---

(Not used.)

 

 

 

 

2

MSG

Message

Interrupts when an advisory has been displayed

 

 

 

on the oscilloscope.

 

 

 

 

1

USR

User Event

Interrupts when enabled user event conditions

 

 

 

occur.

 

 

 

 

0

TRG

Trigger

Interrupts when a trigger occurs.

 

 

 

 

Query Syntax

*SRE?

 

The *SRE? query returns the current value of the Service Request Enable

 

Register.

Return Format

<mask><NL>

 

<mask> ::= sum of all bits that are set, 0,..,255;

 

an integer in NR1 format

See Also

"Introduction to Common (*) Commands" on page 63

 

"*STB (Read Status Byte)" on page 81

 