Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 391

:TRIGger:UART:BURSt

(see page 530)
Command Syntax :TRIGger:UART:BURSt <value>
<value> ::= {OFF | 1 to 4096 in NR1 format}
The :TRIGger:UART:BURSt command selects the burst value (Nth
occurrence of trigger event after idle period) in the range 1 to 4096 or
OFF, for the trigger when in UART mode.
Query Syntax :TRIGger:UART:BURSt?
The :TRIGger:UART:BURSt? query returns the current UART trigger burst
value.
Return Format <value><NL>
<value> ::= {OFF | 1 to 4096 in NR1 format}
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:MODE" on page 319
":TRIGger:UART:IDLE" on page 393
":TRIGger:UART:TYPE" on page 399