350 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
3Commands by Subsystem

:TRIGger:GLITch:LEVel

(see page 530)
Command Syntax :TRIGger:GLITch:LEVel <level_argument>
<level_argument> ::= <level>[, <source>]
<level> ::= .75 x full-scale voltage from center screen in NR3 format
for internal triggers
<level> ::= 2 V with probe attenuation at 1:1 in NR3 format for
external triggers
<source> ::= {CHANnel<n> | EXTernal}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger:GLITch:LEVel command sets the trigger level voltage for the
active pulse width trigger.
Query Syntax :TRIGger:GLITch:LEVel?
The :TRIGger:GLITch:LEVel? query returns the trigger level of the current
pulse width trigger mode. If all channels are off, the query returns
"NONE."
Return Format <level_argument><NL>
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:MODE" on page 319
":TRIGger:GLITch:SOURce" on page 354