5Obsolete and Discontinued Commands

:MEASure:TMAX

(see page 530)

Command Syntax :MEASure:TMAX [<source>]

<source> ::= {CHANnel<n> FUNCtion MATH}

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

 

 

The :MEASure:TMAX command installs a screen measurement and starts

 

 

an X- at- Max- Y measurement on the selected waveform. If the optional

 

 

source is specified, the current source is modified.

 

 

The :MEASure:TMAX command is obsolete and is provided for backward compatibility to

N O T E

 

 

previous oscilloscopes. Use the :MEASure:XMAX command (see page 255) instead.

 

 

 

 

Query Syntax

:MEASure:TMAX? [<source>]

 

 

The :MEASure:TMAX? query returns the horizontal axis value at which the

 

 

maximum vertical value occurs on the current source. If the optional

 

 

source is specified, the current source is modified. If all channels are off,

 

 

the query returns 9.9E+37.

Return Format

<value><NL>

 

 

<value> ::= time at maximum in NR3 format

See Also

"Introduction to :MEASure Commands" on page 218

 

 

":MEASure:TMIN" on page 483

 

 

":MEASure:XMAX" on page 255

 

 

":MEASure:XMIN" on page 256

482

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 482
Image 482
Agilent Technologies 5000 Series manual MEASureTMAX, MEASureTMIN on