Commands by Subsystem 3

:MARKer:Y1Position

(see page 530)

Command Syntax :MARKer:Y1Position <position> [suffix]

<position> ::= Y1 cursor position in NR3 format

<suffix> ::= {mV V dB}

The :MARKer:Y1Position command sets :MARKer:MODE to MANual, sets the Y1 cursor position and moves the Y1 cursor to the specified value.

Query Syntax :MARKer:Y1Position?

The :MARKer:Y1Position? query returns current Y1 cursor position. If the front- panel cursors are off an error is returned. This is functionally equivalent to the obsolete :MEASure:VSTArt command/query

Return Format <position><NL>

<position> ::= Y1 cursor position in NR3 format

See Also "Introduction to :MARKer Commands" on page 204

":MARKer:MODE" on page 205

":MARKer:X1Y1source" on page 207

":MARKer:X2Y2source" on page 209

":MARKer:Y2Position" on page 212

":MEASure:VSTArt" on page 490