:TRIGger:LIN:SYNCbreak

 

(see page 530)

Command Syntax

:TRIGger:LIN:SYNCbreak <value>

 

<value> ::= integer = {11 12 13}

 

The :TRIGger:LIN:SYNCbreak command sets the length of the LIN sync

 

break to be greater than or equal to 11,12, or 13 clock lengths. The sync

 

break is the idle period in the bus activity at the beginning of each packet

 

that distinguishes one information packet from the previous one.

Query Syntax

:TRIGger:LIN:SYNCbreak?

 

The :TRIGger:LIN:STANdard? query returns the current LIN sync break

 

setting.

Return Format

<value><NL>

 

<value> ::= {11 12 13}

See Also

"Introduction to :TRIGger Commands" on page 313

 

":TRIGger:MODE" on page 319

 

":TRIGger:LIN:SIGNal:DEFinition" on page 497

 

":TRIGger:LIN:SOURce" on page 368