Anritsu S251C Power Monitor Menu Option, Mode=Power Monitor Softkeys, Functions and Operations

Models: S251C

1 81
Download 81 pages 12.93 Kb
Page 25
Image 25
Power Monitor Menu (Option 5)

Chapter 2 Functions and Operations

Power Monitor Menu (Option 5)

Selecting POWER MONITOR from the Mode menu causes the soft keys, described below, to be displayed and the corresponding values shown in the message area.

MODE=POWER MONITOR:

SOFTKEYS:

UNITS

REL

OFFSET

ZERO

Figure 2-5. Power Monitor Mode Soft Key Labels

￿UNITS — Toggles between dBm and Watts.

￿REL — Turns relative mode OFF, if currently ON. If relative mode is cur- rently OFF, turns it ON and causes the power level to be measured and saved as the base level. Subsequent measurements are then displayed relative to this saved value. With units of dBm, relative mode displays dBr; with units of Watts, relative mode displays % (percent).

￿OFFSET — Turns Offset OFF, if currently ON. If Offset is currently OFF, turns it ON and opens the Offset parameter for data entry. Press ENTER when data entry is complete.

Offset is the attenuation (in dB) inserted in the line between the DUT and the RF detector. The attenuation is added to the measured input level prior to dis- play.

￿ZERO — Turns Zero OFF, if currently ON. If Zero is currently OFF, this softkey turns it ON and initiates collection of a series of power level samples, which are averaged and saved. This saved value is then subtracted from sub- sequent measurements prior to display.

2 - 10

Page 25
Image 25
Anritsu S251C manual Power Monitor Menu Option, Mode=Power Monitor Softkeys, Functions and Operations