Chapter 3 Getting Started

Chapter 3 Getting Started

Select the Frequency or Distance

Regardless of the calibration method used, the frequency range for the desired measure- ments must be set before calibrating the Site Master. The following procedure selects the frequency range for the calibration.

Step 1. Press the FREQ/DIST key.

Step 2. Press the F1 soft key.

Step 3. Enter the desired frequency using the key pad or the Up/Down arrow key. Step 4. Press ENTER to set F1 to the desired frequency.

Step 5. Press the F2 soft key.

Step 6. Enter the desired frequency using the keypad or the Up/Down arrow key. Step 7. Press ENTER to set F2 to the desired frequency.

Check that the start and stop frequencies displayed match the desired measure- ment range..

Calibration

For accurate results, the Site Master must be calibrated at the ambient temperature before making any measurements. The Site Master must be re-calibrated whenever the setup fre- quency changes, the temperature exceeds the initial calibration temperature or when the test port extension cable is removed or replaced.

The Site Master may be calibrated manually with Open, Short, Load (OSL) calibration components. Two port calibration requires two Load components.

If a Test Port Extension Cable is to be used, the Site Master must be calibrated with the Test Port Extension Cable in place. The Test Port Extension Cable is a phase stable cable and is used as an extension cable on the test port to ensure accurate and repeatable measure- ments. This phase stable cable can be moved and bent while making a measurement with- out causing errors in the measurement.

NOTE: The test port extension cable should have the appropriate connectors for the measurement. Use of additional connector adapters after the test port extension cable can contribute to measurement errors not compensated for during calibration.

The following procedure explains the calibration methods. Refer to Figure 3-2for a calibra- tion setup diagram.