Anritsu S251C manual Symbols, Self Test, Icon, Functions and Operations

Models: S251C

1 81
Download 81 pages 12.93 Kb
Page 30
Image 30
Symbols

Chapter 2 Functions and Operations

Symbols

Table 2-1provides a listing of the symbols used as condition indicators on the LCD display.

Table 2-1.LCD Icon Symbols

Icon

HOLD

￿dx

T

￿

Symbol

Site Master is in Hold for power conservation. To resume sweeping, press the RUN/HOLD key. After 10 minutes without a key press, the Site Master will automatically activate the power conservation mode.

Integrator Failure. Intermittent integrator failure may be caused by inter- ference from another antenna. Persistent integrator failure indicates a need to return the Site Master to the nearest Anritsu service center for repair.

Lock fail indication. Check battery. (If the Site Master fails to lock with a fully charged battery, call your Anritsu Service Center.)

When calibration is performed, the Site Master stores the ambient temper- ature. If the temperature drifts outside the specified range, this indicator will flash. A recalibration at the current temperature is recommended.

Indicates the remaining charge on the battery. The inner white rectangle grows longer as the battery charge depletes.

Indicates internal data processing.

CAL ON The Site Master has been calibrated.

CALL OFF The Site Master has not been calibrated.

Self Test

At turn-on, the Site Master runs through a series of quick checks to ensure the system is functioning properly. Note that the battery voltage and temperature are displayed in the lower left corner below the self test message. If the battery is low, or if the ambient temper- ature is not within the specified operational range, Self Test will fail. If Self Test fails and the battery is fully charged and the Site Master is within the specified operating temperature range, call your Anritsu Service Center.

2 - 15

Page 30
Image 30
Anritsu S251C manual Symbols, Self Test, Icon, Functions and Operations