Figure 1-2. Two-Wire In Situ Oxygen Analyzer HART Connections and AMS Application

 

e. Handling the Analyzer

 

 

 

 

 

 

 

 

 

 

 

 

 

f.

System Considerations

The probe was specially packaged to pre- vent breakage due to handling. Do not re- move the padding material from the probe until immediately before installation.

It is important that printed circuit boards and integrated circuits are handled only when adequate antistatic precautions have been taken to pre- vent possible equipment damage.

The oxygen probe is designed for in- dustrial applications. Treat with care to avoid physical damage. The probe contains components made from ce- ramic, which are susceptible to shock when mishandled. THE WARRANTY DOES NOT COVER DAMAGE FROM MISHANDLING.

Prior to installing your Rosemount Two-Wire In Situ Oxygen Analyzer, make sure you have all the components necessary to make the system installation. Ensure all the com- ponents are properly integrated to make the system functional.

After verifying that you have all the compo- nents, select mounting locations and deter- mine how each component will be placed in terms of available line voltage, ambient temperatures, environmental considera- tions, convenience, and serviceability. Fig- ure 1-2 shows a typical system wiring. A typical system installation is illustrated in Figure 1-3.

Rosemount Analytical Inc. A Division of Emerson Process Management