AIR SUPPLY (REFERENCE AIR)

PRESSURE

FLOWMETER REGULATOR

MODEL 3081

TRANSMITTER

4-20 mA SIGNAL

26020038

Figure 1-3. Typical System Installation

A source of instrument air is required at the oxygen probe for reference air use. Since the Two-Wire In Situ Oxygen Analyzer is equipped with an in-place calibration fea- ture, provisions should be made for con- necting calibration check gas tanks to the oxygen probe during calibration.

If the calibration check gas bottles are to be permanently connected, a check valve is required next to the calibration fittings on the integral electronics.

This check valve is to prevent breathing of calibration check gas line and subsequent flue gas condensation and corrosion. The

check valve is in addition to the stop valve in the calibration check gas kit.

NOTE

The electronics of the Model 3081 Transmitter is rated NEMA 4X (IP65) and is capable of operating at tem- peratures up to 65°C (149°F).

Retain the packaging in which the Rosemount Two-Wire In Situ Oxygen Analyzer arrived from the factory in case any components are to be shipped to another site. This packag- ing has been designed to protect the product.