Model 3081FG
Instruction Manual
CELL+(BK) |
| THERMOCOUPLE+(GY) | |||
|
|
|
| ||
JUMPER |
|
|
|
| |
TERMINAL | 7 | 8 | 9 | 10 | |
BLOCK (TB1) | |||||
6 |
|
| 11 |
| 5 | 12 |
PROBE CABLE | 4 | 13 |
3 | 14 | |
SHIELD | 2 | 15 |
| 1 |
|
GROUND |
| 16 |
EARTH GROUND TERMINALS
b. Reference Air Package
After the oxygen probe is installed, connect
the reference air set. Install the reference air 2 set according to Figure
c. | Instrument Air (Reference Air) |
| Instrument air is required for reference. Use |
| 10 psig (68.95 kPa gage) minimum, 225 |
| psig (1551.38 kPa gage) at 0.2 scfh (100 |
| ml/min.); less than 40 |
| hydrocarbons. Regulator outlet pressure |
| should be set at 5 psi (35 kPa). |
d. | Calibration Check Gas |
| Two calibration check gas concentrations |
CELL AND | |
THERMOCOUPLE |
|
CONDUITS
NOTE: RUN CELL AND THERMOCOUPLE SIGNALS IN SEPARATE CONDUIT FROM
26020005
Figure 2-10. Model 3081 Transmitter Terminal
Block
2-4 PNEUMATIC INSTALLATION
a.General
Reference air is required for O2 calculation, and calibration check gas is required during a calibration check. Refer to Figure
are used with the |
Analyzer: Low Gas - 0.4% O2 and High Gas |
- 8% O2, each with the balance in nitrogen. |
Do not use 100% nitrogen. See Figure |
for the probe connections. Set both calibra- |
tion check gases at the same flow rate: 5 |
scfh (2.5 L/min). |
1/4 TUBE FITTING (REFERENCE AIR PORT)
1/4 TUBE FITTING (CALIBRATION CHECK GAS PORT)
REFERENCE
AIR VENT | 26020006 |
Figure 2-11. Oxygen Probe Gas Connections
Rosemount Analytical Inc. A Division of Emerson Process Management | Installation |