Model 3081FG

Instruction Manual

IB-106-3081 Rev. 1.5 September 2002

FACTORY-INSTALLED

CELL+(BK)

CELL-(WH) THERMOCOUPLE-(RD)

 

THERMOCOUPLE+(GY)

 

 

 

 

JUMPER

 

 

 

 

TERMINAL

7

8

9

10

BLOCK (TB1)

6

 

 

11

 

5

12

PROBE CABLE

4

13

3

14

SHIELD

2

15

 

1

 

GROUND

 

16

EARTH GROUND TERMINALS

4-20 mA (-)

4-20 mA (+)

b. Reference Air Package

After the oxygen probe is installed, connect

the reference air set. Install the reference air 2 set according to Figure 2-12.

c.

Instrument Air (Reference Air)

 

Instrument air is required for reference. Use

 

10 psig (68.95 kPa gage) minimum, 225

 

psig (1551.38 kPa gage) at 0.2 scfh (100

 

ml/min.); less than 40 parts-per-million total

 

hydrocarbons. Regulator outlet pressure

 

should be set at 5 psi (35 kPa).

d.

Calibration Check Gas

 

Two calibration check gas concentrations

CELL AND

4-20 mA

THERMOCOUPLE

 

CONDUITS

NOTE: RUN CELL AND THERMOCOUPLE SIGNALS IN SEPARATE CONDUIT FROM 4-20 mA LINE.

26020005

Figure 2-10. Model 3081 Transmitter Terminal

Block

2-4 PNEUMATIC INSTALLATION

a.General

Reference air is required for O2 calculation, and calibration check gas is required during a calibration check. Refer to Figure 2-11 for the gas connections on the oxygen probe.

are used with the Two-Wire In Situ Oxygen

Analyzer: Low Gas - 0.4% O2 and High Gas

- 8% O2, each with the balance in nitrogen.

Do not use 100% nitrogen. See Figure 2-11

for the probe connections. Set both calibra-

tion check gases at the same flow rate: 5

scfh (2.5 L/min).

1/4 TUBE FITTING (REFERENCE AIR PORT)

1/4 TUBE FITTING (CALIBRATION CHECK GAS PORT)

REFERENCE

AIR VENT

26020006

Figure 2-11. Oxygen Probe Gas Connections

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-11

Page 29
Image 29
Emerson 3081FG Reference Air Package, Instrument Air Reference Air, Calibration Check Gas, Pneumatic Installation