Instruction Manual
Model 3081FG
O2 CONCENTRATION | O2 CONCENTRATION |
%
%
O2 CELL
IMPEDANCE STATUS
26020020
Figure 3-23. CELL IMPEDANCE
d.CELL IMPEDANCE
Refer to Figure
WARN indicates the cell has degraded but is still operational. HI indicates that the cell has degraded but is still operational; how- ever, failure will occur soon. Pressing NEXT accesses the PREVIOUS SLOPE screen.
CALCHECK CELL
SLOPE VALUE
26020052
Figure 3-24. PREVIOUS SLOPE
f.PREVIOUS CONSTANT
Refer to Figure
NOTE
Temperature influences cell imped- ance. Wait until the cell is at operating temperature before checking cell im- pedance. If checked before the cell reaches operating temperature [550°C (1022°F)], this screen displays a fail indication.
e.PREVIOUS SLOPE
Refer to Figure
O2 CONCENTRATION
%
CALCHECK CELL
CONSTANT VALUE
26020053
Figure 3-25. PREVIOUS CONSTANT