Instruction Manual
Model 3081FG
a.IN MANUAL? Refer to Figure
Failure to remove the analyzer from automatic control loops prior to per- forming this procedure may result in a dangerous operating condition.
If the O2 output value is used in any auto- matic process control loops, the loop must be placed in manual before beginning a calibration check.
O2 CONCENTRATION
%
%
HIGH CALIBRATION
CHECK GAS O2 VALUE
Once the analyzer is removed from any automatic control loops, press ENTER to edit the screen. Use the arrow keys to se- lect Y (yes) and press ENTER to process the selection and to display the ACCEPT HIGH O2 screen.
O2 CONCENTRATION
%
Y/N (FOR USER
RESPONSE)
26020054
Figure 3-27. IN MANUAL?
26020055
Figure 3-28. ACCEPT HIGH O2
b.ACCEPT HIGH O2
Refer to Figure
c.ACCEPT LOW O2
Refer to Figure
Rosemount Analytical Inc. A Division of Emerson Process Management |