Figure 6-7. Fault 3, Shorted Thermocouple

c. Fault 3, Shorted Thermocouple

The thermocouple connections are shorted.

The fault displays as shown in Figure 6-7.

1. Allow adequate time for the oxygen

26020027

Figure 6-8. Fault 4, High Probe Temperature

d.Fault 4, High Probe Temperature

The probe’s temperature has exceeded the maximum cell temperature setpoint. The fault displays as shown in Figure 6-8.

probe to reach operating temperatures.

Probe temperatures below approxi-

mately 500°C (932°F) may result in this

fault.

1. If the probe temperature exceeds the maximum cell temperature setpoint, the 4-20 mA signal output will become invalid and go to the default value.

6

2.

Refer to Figure 6-4.Using a multime-

 

ter, measure between terminals 8(-)

 

and 10(+).

3.

If the multimeter reading, in voltage

 

mode, is between -0.5 and +0.5 mV,

 

the thermocouple is shorted.

4.

If the thermocouple is shorted, replace

 

the oxygen probe per paragraph 5-2.

5.

If the thermocouple is not shorted, then

 

replace the PC board stack assembly

 

per paragraph 5-1.