d.Navigation

The IRC in Figure 3-6is used to interact with the Model 3081 Transmitter and navi- gate through the screens on the LCD.

1.Hold the IRC within 6 ft (1.8 m) of the Model 3081 Transmitter and within 15 degrees from the centerline of the transmitter LCD. The amount of ambi- ent light may also affect IRC performance.

NOTE

The LCD may react slowly to IRC commands. Allow sufficient time be- tween key presses to avoid undesired or repeated commands from accumu- lating in the command queue.

2.Use the keys on the IRC to navigate through the menu screens. Refer to Figure 3-6.General usage is as follows:

(a)RESET. Returns to the PROCESS DISPLAY screen at the top of the menu tree. Any non-entered num- ber in the exited state will be ig- nored, and the previous data will be used.

(b)HOLD. Not used.

(c)Left/Right Arrow. Moves left and right among editable digits on the display.

(h)ENTER. Initiates the editing proc- ess and causes the most signifi- cant digit of the edited item to start flashing. Also processes the entry so the previous value updates to the new value entered using the ar- row keys. Failure to press ENTER

before exiting a screen will cancel

 

the input value and revert to the

 

previous value.

3

 

(i)NEXT. Accesses the next user screen as shown in the menu tree. Any non-entered number in the ex- ited state will be ignored, and the previous data will be used.

(j)EXIT. Exits from sub-branches of the menu tree where an exit option is explicitly shown. Otherwise, re- turns to the PROCESS DISPLAY screen at the top of the menu tree. Any non-entered number in the ex- ited state will be ignored, and the previous data will be used.

 

RESET

 

HOLD

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAL

 

ENTER

 

 

 

(d)Up/Down Arrow. Increases or de- creases the value of the currently selected digit on the display.

(e)CAL. Accesses the CALCHECK MENU branch of the menu tree. Only works from the PROCESS DISPLAY screen.

(f)PROG. Accesses the PROGRAM MENU branch of the menu tree. Only works from the PROCESS DISPLAY screen.

(g)DIAG. Accesses the DIAGNOS- TICS MENU branch of the menu tree. Only works from the PROC- ESS DISPLAY screen.

PROG

 

NEXT

 

 

 

 

 

 

DIAG

 

EXIT

 

 

 

MODEL 3081

REMOTE CONTROL

ROSEMOUNT

FISHER-ROSEMOUNT

26020032

Figure 3-6. Infrared Remote Control (IRC)

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup and Operation 3-5

Page 35
Image 35
Emerson 3081FG instruction manual Navigation, Infrared Remote Control IRC