Rosemount 5300/5400 Series

Manual Supplement

00809-0500-4530, Rev AA October 2010

ESTABLISH HART

COMMUNICATION

Connect to the MA/MB terminals

Figure 1-8. RS-485 Communication after startup

The Rosemount 5300 Series and Rosemount 5400 Series can be configured using the Rosemount Radar Master (RRM) PC software or a Field Communicator. Configuration is done by sending HART commands through the HART to Modbus Converter (HMC) to the 5300/5400 transmitter electronics. To establish HART communication, connect to the MA/MB terminals, or to the HART terminals. Both alternatives are described below.

The 5300/5400 level transmitter can be configured with RRM using the MA, MB terminals.

An RS-485 Converter is required to connect to the transmitter.

The transmitter will try to establish communication using different protocols during 20 second timeslots from time of startup.

 

 

 

 

Configured

 

 

 

Configured

 

 

 

 

 

 

protocol

 

 

 

protocol

 

 

 

Modbus RTU

 

HART

 

(Modbus RTU,

 

HART

 

(Modbus RTU,

 

 

20 seconds

 

20 seconds

 

Levelmaster,

 

20 seconds

 

Levelmaster,

 

 

 

 

 

 

or Modbus

 

 

 

or Modbus

 

 

 

 

 

 

 

ASCII)

 

 

 

ASCII)

 

Time

 

 

 

 

20 seconds

 

 

 

20 seconds

 

 

 

 

 

 

 

 

 

 

 

0 s

20 s

40 s

60 s

80 s

100 s

The transmitter will continue to use a communication protocol once communication has been established.