SPECIFICATIONS

Table 1-26. Specifications

Power supply

8-30 Vdc (max. rating)

 

 

Power consumption

< 0.5 W (with HART address=1)

< 1.2 W (incl. four HART slaves)

 

 

 

 

Two-wire half duplex RS-485 Modbus. Use shielded

Signal wiring

twisted pair wiring, preferably with an impedance of 120Ω

(typically 24 AWG), in order to comply with EIA-485

 

 

standard and EMC regulations.

 

 

 

The power supply cables must be suitable for the supply

Power supply cabling

voltage and ambient temperature, and approved for use

 

in hazardous areas, where applicable.

 

 

Ground (common mode)

± 7 V

voltage limit

 

 

 

Bus termination

Standard RS-485 bus termination per EIA-485

 

 

See the Rosemount 5300 Series Reference Manual (Document No. 00809-0100-4530), and the Rosemount 5400 Series Reference Manual (Document No. 00809-0100-4026) for further specifications.