Emerson 5300, 5400 specifications Output Variables for Floating Point Format Code

Models: 5300 5400

1 42
Download 42 pages 48.09 Kb
Page 27
Image 27

Manual Supplement

00809-0500-4530, Rev AA October 2010

Rosemount 5300/5400 Series

Table 1-14. Output Variables for

 

 

 

Floating Point Format Code 0

 

 

 

Register Name

Register Number

Note

 

 

 

 

Bit information in bitfield:

 

 

 

Bit 0: Invalid Measurement Slave 1 PV.

 

 

 

Bit 1: Invalid Measurement Slave 1 SV.

 

 

 

Bit 2: Invalid Measurement Slave 1 TV.

 

 

 

Bit 3: Invalid Measurement Slave 1 FV.

 

Slave 1 Status

2000

Bit 14: HART bus busy (slave in burst or

 

other master present)

 

 

 

 

 

 

Bit 15: HTM Task not running (option not

 

 

 

available).

 

 

 

Note: Bit 1-3 is set when Invalid

 

 

 

Measurement of Slave 1 Non PV, i.e. all

 

 

 

three bits are set simultaneously.

 

 

 

Primary variable from slave 1

 

Slave 1 PV

2002

represented in IEEE 754 format, using

 

 

 

Floating Point Format Code 0.

 

 

 

Secondary variable from slave 1

 

Slave 1 SV

2004

represented in IEEE 754 format, using

 

 

 

Floating Point Format Code 0.

 

 

 

Tertiary variable from slave 1

 

Slave 1 TV

2006

represented in IEEE 754 format, using

 

 

 

Floating Point Format Code 0.

 

 

 

Fourth variable from slave 1 represented

 

Slave 1 FV (QV)

2008

in IEEE 754 format, using Floating Point

 

 

 

Format Code 0.

Table 1-15. Output Variables for

 

 

 

Floating Point Format Code 2

 

 

 

Register Name

Register Number

Note

 

 

 

 

Bit information in bitfield:

 

 

 

Bit 0: Invalid Measurement Slave 1 PV.

 

 

 

Bit 1: Invalid Measurement Slave 1 SV.

 

 

 

Bit 2: Invalid Measurement Slave 1 TV.

 

 

 

Bit 3: Invalid Measurement Slave 1 FV.

 

Slave 1 Status

2100

Bit 14: HART bus busy (slave in burst or

 

other master present)

 

 

 

 

 

 

Bit 15: HTM Task not running (option not

 

 

 

available).

 

 

 

Note: Bit 1-3 is set when Invalid

 

 

 

Measurement of Slave 1 Non PV, i.e. all

 

 

 

three bits are set simultaneously.

 

 

 

Primary variable from slave 1

 

Slave 1 PV

2102

represented in IEEE 754 format, using

 

 

 

Floating Point Format Code 2.

 

 

 

 

1-27

Page 27
Image 27
Emerson 5300, 5400 specifications Output Variables for Floating Point Format Code