Manual Supplement

00809-0500-4530, Rev AA October 2010

Rosemount 5300/5400 Series

To configure the 5300/5400 level transmitter using RRM and the MA, MB terminals, do the following:

1.Connect the RS-485 Converter to the MA, MB connectors.

2.Start RRM and open Communication Preferences.

3.Enable HART communication and make sure the port for the RS-485 Converter is selected. Use the following settings:

4.Connect the power wires (or cycle power) to the transmitter.

5.Wait 20 seconds and then open the Search Device window in RRM (also see note below). Make sure HART address 1 is being scanned.

1-11

Page 11
Image 11
Emerson specifications Rosemount 5300/5400 Series