Instruction Manual

IB-106-340CDR Original Issue January, 2002

Hazardous Area Oxymitter DR

1-4 SPECIFICATIONS

 

Hazardous Area Certifications

CENELEC EEXd IIB + H2 T3

 

CSA Class I, Division 1, Groups B, C, D

Probe Lengths

18 in. (457 mm)

 

3 ft (0.91 m)

 

6 ft (1.83 m)

Temperature Limits in Process

 

Measurement Area

0° to 704°C (32° to 1300°F)

 

up to 1300°C (2400°F) with optional accessories

Resolution Sensitivity

0.01% O2 transmitted signal

Sensing Cell Repeatability

±0.75% of O2 reading, or 0.05% O2

System Response to Calibration Gas

Initial response in less than 3 seconds T90 in less than

 

8 seconds

Resolution Sensitivity

0.01% of O2 value

Mounting and Mounting Position

Vertical or horizontal

Materials:

 

Probe

Wetted or welded parts - 316L stainless steel

 

Non-wetted parts - 304 stainless steel, low-copper

 

aluminum

Termination Housing

Low-copper aluminum

Calibration Gas Mixtures Recommended

0.4% O2, Balance N2

 

8% O2, Balance N2

Calibration Gas Flow

2.5 l/m (5 scfh)

Optional Reference Air

1 l/m (2 scfh), clean, dry, instrument-quality air

 

(20.95% O2), regulated to 34 kPa (5 psi), Optional

Heater Voltage

115 ±10% VAC, 50/60 Hz., 200VA

Thermocouple

Type K

Power Requirements:

 

Nominal

175 W

Maximum

500 W

Ambient Operating Temperature

 

(Junction Box)

93°C (200°F) [71°C (160°F) max

 

for YEW replacement]

Fisher-Rosemount has satisfied all obligations coming from the European legislation to harmonize the product requirements in Europe.

1-6 Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 20
Image 20
Emerson ib-106-340cdr, oxymitterdr hazardous area in-siu oxygen probe instruction manual Specifications