Instruction Manual
IB-106-340CDR Original Issue
January, 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-15

Hazardous Area Oxymitter DR
f. WORLD CLASS 3000 INTELLIGENTFIELD TRANSMITTER SETUPThe DR probe operates with a 115 VACheater. Ensure that the voltage selectionjumpers in the IFT or HPS, if used, are setproperly. Refer to Figure 2-13 for IFTjumper selection, and Figure 2-14 for HPSjumper selection. For additional setup andconfiguration information, refer to IB-106-300NFX.
ALWAYSDISCONNECTLINE VOLTAGE
FROMINTELLIGENT FIELD TRANSMITTER
BEFORECHANGING JUMPERS.
JUMPER
CONFIGURATION
LINEVOLTAGE
SELECTION
JUMPER
(INSTALL)
PROBEHEATER
VOLTAGESELECTION
JUMPER
(INSTALL)
100V.A.C.
120V.A.C.
220V.A.C.
240V.A.C.
JM3,JM7, JM2
JM8,JM7, JM1
JM6,JM5, JM2
JM6,JM5, JM1
WORLDCLASS PROBE (44V)
WORLDCLASS "DIRECT
REPLACEMENT" PROBE (115V)
OROXYMITTER DIRECT
REPLACEMENTPROBE (115V)
218PROBE (115V)
JM10
JM9
JM9
36210015
Figure 2-13. IFT 3000 Power Supply Board Jumper ConfigurationIf incorrect heater voltage is selected, damage to the Oxymitter DR may occur. Ref er toIB-106-300NFX for additional configuration information. Always update the relevant labelingto reflect the set voltage.
ALWAYSDISCONNECT LINE VOLTAGE
FROM HEATER POWER SUPPLY AND
ANALOG ELECTRONICS (IF USED)
BEFORE CHANGING JUMPERS.
JUMPER
CONFIGURATIONS
LINE VOLTAGE
SELECTION
JUMPER
(INSTALL)
JUMPER
(INSTALL)
HEATER
POWER JUMPER
100/120V.A.C.
220/240V.A.C.
JM4,JM1
JM5
REMOTE
ON
REMOVE JM2
*INSTALL JM2
ELECTRONICS
SELECTION JUMPER
PROBE HEATER
VOLTAGE SELECTION
INSTALL JM3, JM6
REMOVE JM3, JM6
*ANALOG (EXISTING)
DIGITAL
(NEXT GENERATION)
WORLD CLASS PROBE
(44V)
JM7
JM8
218 PROBE (115V)
1
NOTES:
100 V.A.C. OPERATIONREQUIRES TRANSFORMER PART
NUMBER 1M02961G02.
1
36210014
JM8
DIRECT REPLACEMENT
WORLD CLASS OR
DR OXYMITTER
Figure 2-14. Heater Power Supply (HPS 3000) Jumper Configuration
2