Hazardous Area Oxymitter DR

Instruction Manual

IB-106-340CDR Original Issue January, 2002

f.WORLD CLASS 3000 INTELLIGENT FIELD TRANSMITTER SETUP

The DR probe operates with a 115 VAC heater. Ensure that the voltage selection jumpers in the IFT or HPS, if used, are set

properly. Refer to Figure 2-13for IFT jumper selection, and Figure 2-14for HPS jumper selection. For additional setup and

configuration information, refer to IB-106- 2 300NFX.

JUMPER

CONFIGURATION

ALWAYS DISCONNECT LINE VOLTAGE FROM INTELLIGENT FIELD TRANSMITTER BEFORE CHANGING JUMPERS.

LINE VOLTAGE

JUMPER

SELECTION

(INSTALL)

 

 

100 V.A.C.

JM3, JM7, JM2

 

 

120 V.A.C.

JM8, JM7, JM1

 

 

220 V.A.C.

JM6, JM5, JM2

 

 

240 V.A.C.

JM6, JM5, JM1

 

 

PROBE HEATER

JUMPER

VOLTAGE SELECTION

(INSTALL)

 

 

WORLD CLASS PROBE (44V)

JM10

 

 

218 PROBE (115V)

JM9

 

 

WORLD CLASS "DIRECT

JM9

REPLACEMENT" PROBE (115V)

 

OR OXYMITTER DIRECT

 

REPLACEMENT PROBE (115V)

 

 

 

36210015

Figure 2-13. IFT 3000 Power Supply Board Jumper Configuration

If incorrect heater voltage is selected, damage to the Oxymitter DR may occur. Refer to IB-106-300NFX for additional configuration information. Always update the relevant labeling to reflect the set voltage.

 

 

 

 

 

 

 

ALWAYS DISCONNECT LINE VOLTAGE

 

 

 

 

 

 

 

 

FROM HEATER POWER SUPPLY AND

 

 

 

 

 

 

 

 

 

 

 

JUMPER

 

ANALOG ELECTRONICS (IF USED)

 

 

 

 

CONFIGURATIONS

BEFORE CHANGING JUMPERS.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LINE VOLTAGE

JUMPER

 

 

 

HEATER

 

JUMPER

 

 

 

SELECTION

(INSTALL)

 

 

 

POWER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

100/120 V.A.C.

JM4, JM1

 

 

 

REMOTE

 

REMOVE JM2

 

 

 

220/240 V.A.C.

JM5

 

 

 

ON

 

*INSTALL JM2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PROBE HEATER

JUMPER

 

 

ELECTRONICS

 

JUMPER

 

 

 

VOLTAGE SELECTION

(INSTALL)

 

 

 

SELECTION

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WORLD CLASS PROBE

JM7

 

 

*ANALOG (EXISTING)

 

INSTALL JM3, JM6

 

 

 

(44V)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIGITAL

 

REMOVE JM3, JM6

 

 

 

 

 

 

 

 

 

 

 

 

 

218 PROBE (115V)

JM8

 

 

(NEXT GENERATION)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIRECT REPLACEMENT

JM8

 

 

 

 

 

 

 

 

 

WORLD CLASS OR

 

 

 

 

 

 

 

 

 

 

 

DR OXYMITTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTES:

 

 

 

 

 

 

 

 

 

 

 

1 100 V.A.C. OPERATION REQUIRES TRANSFORMER PART

 

 

 

 

NUMBER 1M02961G02.

 

 

 

36210014

Figure 2-14. Heater Power Supply (HPS 3000) Jumper Configuration

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-15