Emerson oxymitterdr hazardous area in-siu oxygen probe Heater Strut Assembly, Cell Replacement

Models: ib-106-340cdr oxymitterdr hazardous area in-siu oxygen probe

1 66
Download 66 pages 8.96 Kb
Page 45
Image 45

Hazardous Area Oxymitter DR

Instruction Manual

IB-106-340CDR Original Issue January, 2002

WIRE

LOOP

V-DEFLECTOR

CERAMIC SUPPORT ROD

CERAMIC

DIFFUSER CELL FLANGE

ASSEMBLY

4

HEATER

 

 

27540007

 

Figure 4-3. Heater Strut Assembly

f.

Grasp the wire loop and carefully slide the

structions in paragraph 4-5steps c

 

strut out of the probe tube (Figure 4-3).

through e.

g.

When replacing the strut, align the slot on

l. Follow the instructions in paragraph 4-3bto

 

the heater plate with the calibration gas line

install the Hazardous Area Oxymitter DR

 

in the probe tube. Slide the strut into the

into the stack or duct.

 

probe tube. It will turn to align the hole on

 

 

the back plate of the strut with the calibra-

4-7 CELL REPLACEMENT

 

tion gas line. When the hole and the cali-

 

 

bration gas line are aligned correctly, the

This paragraph covers oxygen sensing cell re-

 

strut will slide in the rest of the way.

placement. Do not attempt to replace the cell

h.

Push down on the back plate of the strut to

until all other possibilities for poor performance

have been considered. If cell replacement is

 

make sure you have spring tension and

needed, order the cell replacement kit

 

then install the pressure clamp (28) on the

(Table 7-1).

 

back plate.

The cell replacement kit (Figure 4-4)contains a

i.

Install capillary breather tubes (23) and tube

cell and flange assembly, corrugated seal,

 

nuts (26) on the CAL, REF, and VENT

setscrews, socket head cap screws, and anti-

 

ports.

seize compound. The items are carefully pack-

j.

 

aged to preserve precise surface finishes. Do

Replace the CAL and REF gas silicon

not remove items from the packaging until they

 

tubes.

are ready to be used. Spanner wrenches and

 

 

hex wrenches needed for this procedure are

k.

Install the termination housing per the in-

part of an available special tools kit (Table 7-1).

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 4-5

Page 45
Image 45
Emerson oxymitterdr hazardous area in-siu oxygen probe, ib-106-340cdr Heater Strut Assembly, Cell Replacement