Table 5-1. Fault Finding (Continued)

 

SYMPTOM

CHECK

FAULT

REMEDY

 

Heater is hot and

Recorder chart

Recorder failure

See Recorder Instruction

 

T/C mV output is at

 

 

Manual.

 

setpoint ±0.2 mV

 

 

 

 

 

Cell mV input to elec-

No cell mV at probe

Replace cell or return

 

 

tronics and cell mV at

when test gas applied

probe to Rosemount.

 

 

probe head

 

 

 

 

 

Probe cell mV OK but no

Check out cable con-

 

 

 

input to electronics

nection.

 

 

 

Cell mV satisfactory both

Check electronics pack-

 

 

 

at junction box and input

age.

 

 

 

to electronics - failure of

 

 

 

 

electronics

 

2.

System responds to

 

 

 

 

oxygen concentra-

 

 

 

 

tion changes but has

 

 

 

 

incorrect indication

 

 

 

 

Good response, with

Recorder or remote

Calibration error

Recalibrate recorder or

 

 

incorrect indication

indicator

 

indicator. Reference Re-

 

 

 

 

corder Instruction Man-

 

 

 

 

ual.

 

 

System calibration

Calibration error

Recalibrate system. Re-

 

 

 

 

place cell if necessary.

 

 

Probe mounting and

Air ingress into duct

Stop air leaks or resite

 

 

condition of duct

 

probe.

 

 

Cell mV input to elec-

Failure of electronics

Check electronics pack-

 

 

tronics

 

age.

3. Probe does not give

Test gas input port

Blocked port

Clean port. If the flue

 

accurate indicationof

 

 

gas is condensing in the

 

applied test gas

 

 

test gas line, insulate the

 

 

 

 

back of the probe. Make

 

 

 

 

sure that the test gas

 

 

 

 

line is capped between

 

 

 

 

calibrations, or a check

 

 

 

 

valve is installed.

 

 

Ceramic diffusion ele-

Diffusion element

Replace diffusion ele-

 

 

ment broken

cracked, broken, or

ment.

 

 

 

missing

 

4. Probe takes a long

Plugged diffusion ele-

Plugged diffusion ele-

Change diffusion ele-

 

time to return to flue

ment

ment

ment or snubber diffu-

 

gas value after

 

 

sion element.

 

calibration gas is

 

 

 

 

turned off

 

 

 

5.

Probe heater

Proper voltage heater is

Wrong heater

Change heater to proper

 

temperature unstable

installed

 

voltage.

5-2 Troubleshooting