Instruction Manual
Hazardous Area Oxymitter DR
Table
| SYMPTOM | CHECK | FAULT | REMEDY |
| Heater is hot and | Recorder chart | Recorder failure | See Recorder Instruction |
| T/C mV output is at |
|
| Manual. |
| setpoint ±0.2 mV |
|
|
|
|
| Cell mV input to elec- | No cell mV at probe | Replace cell or return |
|
| tronics and cell mV at | when test gas applied | probe to Rosemount. |
|
| probe head |
|
|
|
|
| Probe cell mV OK but no | Check out cable con- |
|
|
| input to electronics | nection. |
|
|
| Cell mV satisfactory both | Check electronics pack- |
|
|
| at junction box and input | age. |
|
|
| to electronics - failure of |
|
|
|
| electronics |
|
2. | System responds to |
|
|
|
| oxygen concentra- |
|
|
|
| tion changes but has |
|
|
|
| incorrect indication |
|
|
|
| Good response, with | Recorder or remote | Calibration error | Recalibrate recorder or |
| ||||
| incorrect indication | indicator |
| indicator. Reference Re- |
|
|
|
| corder Instruction Man- |
|
|
|
| ual. |
|
| System calibration | Calibration error | Recalibrate system. Re- |
|
|
|
| place cell if necessary. |
|
| Probe mounting and | Air ingress into duct | Stop air leaks or resite |
|
| condition of duct |
| probe. |
|
| Cell mV input to elec- | Failure of electronics | Check electronics pack- |
|
| tronics |
| age. |
3. Probe does not give | Test gas input port | Blocked port | Clean port. If the flue | |
| accurate indicationof |
|
| gas is condensing in the |
| applied test gas |
|
| test gas line, insulate the |
|
|
|
| back of the probe. Make |
|
|
|
| sure that the test gas |
|
|
|
| line is capped between |
|
|
|
| calibrations, or a check |
|
|
|
| valve is installed. |
|
| Ceramic diffusion ele- | Diffusion element | Replace diffusion ele- |
|
| ment broken | cracked, broken, or | ment. |
|
|
| missing |
|
4. Probe takes a long | Plugged diffusion ele- | Plugged diffusion ele- | Change diffusion ele- | |
| time to return to flue | ment | ment | ment or snubber diffu- |
| gas value after |
|
| sion element. |
| calibration gas is |
|
|
|
| turned off |
|
|
|
5. | Probe heater | Proper voltage heater is | Wrong heater | Change heater to proper |
| temperature unstable | installed |
| voltage. |