Emerson ib-106-340cdr, oxymitterdr hazardous area in-siu oxygen probe Analog Electronics, Setup

Models: ib-106-340cdr oxymitterdr hazardous area in-siu oxygen probe

1 66
Download 66 pages 8.96 Kb
Page 32
Image 32

Instruction Manual

IB-106-340CDR Original Issue January, 2002

Hazardous Area Oxymitter DR

b. WESTINGHOUSE MODELS 218, 225,

5. Adjust potentiometer M110-1 to read -

AND 132 (ANALOG) ELECTRONICS

322.3 millivolts nominal.

SETUP

 

Before beginning operation, it is important

6.

Remove voltmeter leads.

 

 

 

 

 

 

 

 

 

that the probe heater setpoint of the existing

7.

Remove jumper wire.

 

electronics be changed to support the Haz-

 

 

 

 

 

 

 

 

ardous Area Oxymitter DR In-Situ Oxygen

 

 

 

 

 

 

 

 

Probe. The setpoint adjustment procedure

 

TP1

 

M110-1

 

 

required for Models 218, 225, and 132

 

 

 

 

 

 

 

 

analog electronics is as follows:

 

 

 

 

 

 

 

 

1. Open electronics enclosure.

 

 

 

 

 

 

 

 

2. On temperature controller card, Figure

 

 

 

 

 

 

 

 

2-9,connect jumper wire from TP3 to

 

 

 

 

 

 

 

 

either Pin 2 or Pin 7.

 

 

 

 

 

 

 

 

3. Set voltmeter to read DC millivolts

TP3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(MV).

 

 

 

 

 

 

 

 

4. Attach voltmeter with positive (+) lead

 

 

 

 

 

 

 

 

on TP1 and negative (-) on either Pin 2

 

 

 

 

 

 

 

 

or 7.

 

 

 

 

 

 

 

 

NOTE

 

 

 

 

 

 

 

 

The voltage given above is for an am-

PIN 2

2

3

5

6

7

10 12

 

(PSC)

 

 

 

 

 

 

 

bient (machinery space) temperature

 

 

 

 

 

 

 

 

of 25°C (77°F). For each degree of am-

 

 

 

 

 

 

 

 

bient temperature above or below 25°C

 

 

 

 

 

 

 

 

(77°F), add or subtract 0.242 mV from

 

 

 

 

 

 

 

 

the nominal. Example: at 31°C (87°F),

 

 

 

 

PIN 7 (PSC)

P0019

the nominal voltage of -322.3 Mv

Figure 2-9. Temperature Controller Card

should be increased (made less nega-

tive) by 10 x 0.242 or 2.42 mV, making

 

Calibration Points

 

the adjusted nominal -319.9 Mv.

 

 

 

 

 

 

 

 

2-10 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 32
Image 32
Emerson ib-106-340cdr, oxymitterdr hazardous area in-siu oxygen probe instruction manual Analog Electronics, Setup