Instruction Manual

IB-106-340CDR Original Issue January, 2002

Hazardous Area Oxymitter DR

g.THE YOKOGAWA ZA8C AND AV8C CONVERTER ELECTRONICS SETUP

The Hazardous Area Oxymitter DR probe can be wired to work with the Yokogawa®

ZA8C and AV8C Converters. Connect the cabling from the ZA8C or AV8C terminal to the probe terminal in the junction box as shown in Figure 2-15.

 

 

 

 

ZA8C

 

 

 

 

 

 

 

 

OR

 

 

 

 

 

 

 

 

AV8C

 

 

 

 

 

 

 

TERMINAL

 

 

 

 

 

 

 

}

3+

 

 

 

 

 

 

CELL

 

 

 

 

 

 

 

 

4–

 

 

COLD

 

 

 

}

 

 

 

 

 

 

 

 

JUNCTION

 

 

 

5+

 

 

DEVICE

 

THERMOCOUPLE

 

 

 

+

 

 

 

6–

 

 

 

 

 

}

 

 

 

 

 

 

COLD

7+

 

 

 

 

 

 

 

 

 

 

 

 

 

JUNCTION

8–

 

 

 

 

 

 

 

 

 

 

 

 

32

GND

PROBE

16

 

 

 

 

 

 

 

 

 

SHIELD

 

 

 

 

 

HEATER

 

 

 

 

 

 

17

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+

 

+

 

ZA8C OR AV8C

 

 

-

-

 

 

 

 

 

 

 

 

ZA8C

OR

AV8C

CONVERTER

HEATER AND SIGNAL CABLE

SHIELD

STACK WALL

EXISTING

MOUNTING

OXYMITTER DR PROBE

WITH ADAPTER FLANGE

NOTES: 1. HEATER TEMPERATURE SET TO 1380oF (750oC)

2.THE GREATER MASS OF THE OXYMITTER DIRECT REPLACEMENT PROBE REQUIRES LONGER TIME TO HEAT UP. UPON STARTUP, THE YOKOGAWA ELECTRONICS MAY INDICATE AN ERROR BECAUSE THE PROBE HAS NOT REACHED TEMPERATURE SETPOINT IN THE NORMAL TIME. REMOVE POWER FROM THE YOKOGAWA ELECTRONICS OR PROBE

MODULE TO CLEAR THE ERROR, AND RESTORE POWER. THIS PROCEDURE MAY HAVE TO

 

BE REPEATED A COUPLE OF TIMES BEFORE PROBE OPERATING TEMPERATURE IS REACHED.

36210009

Figure 2-15. DR Probe Wired to the ZA8C or A V8C Converter

2-16 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 38
Image 38
Emerson ib-106-340cdr Yokogawa ZA8C and AV8C Converter Electronics Setup, DR Probe Wired to the ZA8C or a V8C Converter