Chapter 10 Configuring your instrument by using wizards

Page 10.13

10.12 Line density wizard

This wizard configures the 795x for getting line density ‘A’ and/or line density ‘B’ data.

"Temp
and
VOS"
"No"
"Yes"
"VOS" Appears for
"Fallback value"
only
"No"
"None"
"Temp"
"Auto" Follow
"A"+"B"
route
"B"
"A"
Transducer A K0
Edit
Line density A calc ?
No
Yes
Transducer A K18
Line density
prime selection ?
A
B
Automatic
"Yes"
A1
Edit Density Limits &
Fallback ?
No
Yes
Line density hi lmt
Exit
wizard
Last Good Value
Fallback value
Prime dens FB type
(Selection)
Line density lo lmt
Transducer A K1
Transducer A K2
None
Density A correction
(Selection)
Temp and VOS
Temp
VOS
Transducer A K19
Line density comp lmt
'b' key
Setup wizard
Line Density
Follow "Temp"
route and then
follow "VOS"
route
A1
Density A offset
Line dens A VOS type
(Selection)
P method
SG method
Density Transducer A
(Selection)
7812 N2 high
7812 N2 low
7811 AR high
7811 N2 high
7811 N2 low
7810 CH4 medium
7810 N2 low
7812 AR high
Line density A gamma
Transducer A K5
Transducer A K6
Density A offset
A2
A1
Transducer A K3
Transducer A K4
Prime density FB val
Appears for
"P method"
only
Appears for
"SG method"
only

“Line density” Wizard Map - Part 1 of 3