Chapter 10 Configuring your instrument by using wizards
Page 10.18
10.14 Specific gravity wizard
This wizard configures the 795x for getting specific gravity ‘A’ and/or specific gravity ‘B’ data.
"No"
"Yes"
"Analogue input"
"Base Density"
"Time Period"
"Auto"
"B"
"A"
"Yes"
"No"
'b' key
Setup wizard
Specific Gravity
Specific Gravity
prime selection ?
A
B
Automatic
Prime SG FB type
(selection)
Fallback value
Last Good Value
Edit SG A calc ?
No
Yes
Edit SG Limits &
Fallback ?
No
Yes
Edit SG B selection ?
Time Period
Base density
Analogue input
"A" route
followed by
the "B" route
I2

I1

Appears for
"Fallback value"
only
Exit
wizard
Prime density FB val
Turn to
"Part 2"
page
SG hi limit
SG low limit
SG comp. limit
SG B K2
SG B K0
Prime Base Density
Base density of Air
SG A K2
SG A K0
From
"Part 2"
page
“Base density” Wizard Map - Part 1 of 2