XDT User’s Manual

3.2 Instrument Technical Specifications

Enclosure:

XDT-OEM bare board

 

 

XDT-NEMA tabletop or surface mount, watertight IP65 tested, and NEMA 12 tested

 

XDT-PM panel-mount, DIN 43700 dimensional standard, optional NEMA 12 type protection gasketing

 

available

 

 

 

Dimensions & Weight:

XDT-OEM: 4.25” x 5.15” x 2.25”, 0.6 lbs (fully optioned) without cables.

 

XDT-NEMA: 4.73” x 6.3” x 3.55”, 2.1 lbs (fully optioned) w/cables.

 

XDT-PM: 5.67” x 2.84” x 2.95” DIN 43700 standard, 1.3 lbs (fully optioned) w/cables.

 

refer to appendix E for detailed dimensions

Environmental Range:

Operating temperature of electronics: 14°F to 122°F (-10°C to 50°C)

 

Storage temperature of electronics: -40°F to 176°F (-40°C to 80°C)

 

Humidity: 0 to 90% RH non-condensing

 

Altitude: 0 to 6500 feet (2000 meters)

Mains Supply Voltage:

100 to 250VAC autoranging, 50/60Hz, 10VA, internal 0.5A, 250V fast acting fuse.

 

Optionally the instrument may be configured with DC power capability: 15 to 30VDC 0.5A.

Electrical connections:

21 contact pluggable screw terminal block, 1.3mm diameter pins on 5mm centers.

Wiring requirements:

For AC Power: 18AWG or heavier wire, an external means for disconnecting the power source is

 

required to meet National Electrical Code requirements.

 

For Sensor signal: RG58 coaxial cable, or RG6 for cable longer than 100’. A 6 foot cable is provided as

 

a standard, other lengths ordered as options. When changing cable, refer to cable compensation section.

Input resolution:

0.1°C dewpoint.

 

 

 

Indicators:

3.5 digit backlit LCD with custom legends, audible indicator.

Engineering units:

°C,°F, PPM, LBS H

2

O/mm scf,gm H O/m3.

 

 

 

2

Controls:

Four push buttons, user’s selections are stored in EEPROM.

Outputs:

Analog: voltage by order or current user selectable 4-20mA or 0/24mA. Linear to the selected engi-

 

neering units, the range is programmable. Output resolution is 0.1°C dewpoint, linearity 1%, max load

 

resistance 500 Ohms.

 

 

Digital RS-232C (9600,8,E,1), can interface to a Personal Computer or other RS-232 device.

Alarm relay contacts:

Ordinary use, explosion-proof housed, and safe area instruments: 10A, 250VAC or 30VDC.

 

Intrinsically Safe Div 2 instruments use hermetically sealed relays: 3A, 120VAC

Isolation:

Sensor is isolated from power ground, analog output and RS-232, but they are shunted with a 33V tran-

 

sorb, a 1M Ohm resistor and 0.1uF capacitor. Refer to Electrical Connections section.

Approvals/Classifications: CE for electromagnetic compatibility, accredited laboratory tested and certified

 

UL and cUL for ordinary field wired use

 

Intrinsically Safe configurations:

for NEC standard IS installations UL,

 

 

 

 

for CENELEC standard IS installations DEMKO,

 

 

 

 

refer to instrument labeling

10