Epson XDT manual display is dark, Legend  denotes a beeping accompaniment to the message

Models: XDT

1 70
Download 70 pages 48.04 Kb
Page 37
Image 37

XDT User’s Manual

 

DISPLAY

EXPLANATION

 

 

REQUIRED ACTION

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LO

ERR



PROM check sum failed.

tests

 

 

 

 

 

 

 

 

HI

ERR



RAM write/read test failed.

 

cycle power

on

 

 

 

 

 

 

 

%

ERR



Unidentified power-up failure.

power

 

if problem persists,

 

 

 

 

 

 

C.S.F/



EEPROM Check Sum Failed.

 

return to your representative

 

 

 

 

 

 

 

 

ERR/AD



A/D converter failure.

2 min.

 

for service.

 

 

 

 

 

ERR/REF



Reference voltage for A/D out of spec.

 

 

 

 

 

 

per

 

 

LO/BAT



Low supply voltage.

 

 

once

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LO RNG/



Instrument low temperature range

tests

 

 

TEMP

 

has been exceeded.

 

make sure that the unit is at a

 

system

 

 

 

 

 

 

 

HI

RNG/



Instrument high temperature range

 

temperature of -10°C to +50°C.

TEMP

 

has been exceeded.

 

 

 

 

 

 

 

 

 

OPN/



Sensor circuit is open.

dewpoint display

 

 

 

 

 

 

 

SAT/



Sensor is saturated.

tests

see table in sensor troubleshooting section 2.5

 

 

 

 

SHR/



Sensor circuit is shorted.

 

 

 

 

 

 

...

 

 

Trying to calculate dewpoint for undefined sensor.

 

select sensor and recalibrate.

 

 

 

 

 

 

 

 

 

 

 

DB

 



Trying to calibrate an undefined sensor.

calibration

 

 

 

 

 

 

tests

see calibration instructions.

SEN/2LO



Sensor reading is ‘too’ low to be from a saturated sensor,

 

 

 

 

thus can not be calibrated

 

 

 

 

 

 

 

 

2LO

 

 

While Compensating for cable length, the measurement did not

cable

tests

Check the means of opening & shorting the sensor

 

 

 

correspond to the expected open or short values.

cable. see section 3.4.5 - 11

2HI

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ERR/EEP



EEPROM write cycle not completed.

 

 

if this persists, return for service.

 

 

 

 

 

 

 

LOC

 



Attempting to modify a locked unit.

 

 

unlock unit, see set-up state 3.4.5.-10.

 

 

 

 

 

 

 

SEC

 



Attempting to unlock a secured unit.

 

 

 

 

 

 

 

 

 

 

HI

(flashing

while

The HI Alarm (alarm #1) relay is deenergized.

 

 

See section 3.4.4.2 Setting the Alarms.

viewing DEWPOINT)

 

 

 

 

 

 

 

 

 

 

 

LO

(flashing

while

The LO Alarm (alarm #2) relay is deenergized.

 

 

 

viewing DEWPOINT)

 

miscellaneous

 

 

 

 

 

 

 

A.O./OPN



The Analog Output ‘sees’ an open circuit

 

check the cables connecting the analog output, make

 

 

 

 

 

sure the terminating impedance is less that 500Ω.

 

 

 

 

 

 

PSI flashing when dis-

Pressure correction is enabled, the dewpoint is computed using

 

refer to section 3.4.4.1

 

 

playing dewpoint

 

programmed pressures

 

 

 

 

 

 

 

 

-70°C (-94°F) with

An XTR-65 sensor is measuring less than -70°C or -94°F D.P.

 

 

check instrument

flashing DEWPOINT

 

 

 

upgrade to XTR-100 sensor

 

 

 

 

 

 

RNG

 

number can not be displayed in 3½ digits

 

 

 

 

 

 

 

 

 

 

XEN

 



Turn on message, Xentaur (Greek Ξ=X)

 

 

 

 

 

 

 

 

display is dark

No power to instrument

 

 

check power to instrument, check fuse

 

 

 

 

 

 

 

Legend:  denotes a beeping accompaniment to the message.

/denotes alternately flashing messages.

30

Page 37
Image 37
Epson XDT display is dark, Legend  denotes a beeping accompaniment to the message, denotes alternately flashing messages