Epson XDT manual b. SpanCheck the sensor c. Dry-down the sensor for at least 12 hrs

Models: XDT

1 70
Download 70 pages 48.04 Kb
Page 33
Image 33

XDT User’s Manual

CAUTION:

2), 3), 4) and 5) are set at the factory and need only be modified when a sensor is changed.

2. Selecting the sensor type: In this mode the user can select the software matching the type of

sensor installed in the instrument: XT1 : XTR-100 (-100°C to +20°C); XT6 : XTR-65 (-65°C to +20°C);

DB : show impedance of any sensor, disable Open and Short error massages.

3&4Adjusting low end sensor attenuation and dewpoint: These modes are used to enter a data pair representing a low dewpoint and the sensor attenuation measured at this low dewpoint. (see Sensor Theory of Operation appendix F). This data pair should not be modified unless the sensor is replaced (refer to procedures for replacing a sensor in appendix L). The sensor ship- ping container is labeled with the proper values.

5.Calibration Adjustment: In this mode the user can enter a sensor specific Adjustment Value to improve the instruments’ accuracy in the range of -50°C to -10°C, this adjustment has very little or no effect outside this range. The Calibration Adjustment Value is printed on the ship- ping container supplied with the sensor, or may be derived if the sensor can be exposed to a known reference.

Deriving an Adjustment Value: If the adjustment value for your sensor is not known, contact your representative and follow the instructions below for entering known values. If the value is not available, then it may be derived using a manual method, provided that a moisture cali- bration facility is available.

The calibration must be done in the range of -25°C to -20°C (attempting to calibrate outside this range may cause inaccuracies), the dewpoint must be kept stable during calibration and it must be measured accurately by a reference instrument such as a chilled mirror. One must also note the state of pressure compensation and turn it on or off as applicable to the reference. The necessary steps are described below (also refer to the Set-Up State flow diagram in Appendix B):

a.Make sure that the low attenuation and low dewpoint are correctly entered for this sensor.

b.SpanCheck™ the sensor.

c.Dry-down the sensor for at least 12 hrs.

d.Expose the sensor to the known dewpoint (-25°C to -20°C), for a sufficient time (at least 30 min.)

e.Go to the CAL/ADJ mode and press the UP button, the display will show the current value, in the range of -1.99 to +1.99 followed by blinking horizontal lines.

f.Observe the display, if the buttons are not being pressed; the display will every few seconds show for a short duration the dewpoint as being computed at the moment (utilizing the current cal-adjust value in the computation).

g.Use the UP or DOWN buttons to modify the cal-adjust value and observe the alternately displayed dewpoint. Perform this adjustment until the dewpoint matches the value shown by the reference instrument.

h.Press the MODE button to go to the next mode, this will save the new cal-adjust value, and it will be retained even if power is turned off.

Entering a known Adjustment Value: If the sensor is being replaced an adjustment value should be supplied with the sensor, to enter it into the instrument:

26

Page 33
Image 33
Epson XDT manual b. SpanCheck the sensor c. Dry-down the sensor for at least 12 hrs