Appendix C

Error Messages

 

The following table lists error message and possible cause associated with

 

 

the HP E1459A Module.

 

 

 

 

 

 

Error

Description

 

 

 

 

 

 

 

 

 

 

-104, “Data type error”

<mask> is not in decimal format.

 

 

 

 

 

 

-109, “Missing parameter”

A command does not include a required parameter (parameter cannot be

 

 

 

defaulted). For example, SENSe:EVENt:PORT0:PEDGe:ENAB - this command

 

 

 

is missing the value of <mask>.

 

 

 

 

 

 

-113, Undefined Header”

An attempt to execute an unrecognized command

 

 

 

 

 

 

-123, “Numeric overflow”

Data for a parameter is out of range. For example, the positive edge detect

 

 

 

enable mask parameter is less than -32768 or greater than +32767.

 

 

 

 

 

 

131, “Unrecognized suffix”

Incorrect suffix such as SECONDS instead of SEC.

 

 

 

 

 

 

-141, “Invalid character data”

Parameter specified is not EXTernal, INTernal, MINimum, MAXimum, DEFault, 0,

 

 

 

1, ON, OFF, DAV, NEDG, PEDG, etc.

 

 

 

 

 

 

-221, “Settings conflict”

Clock source on a port is INTernal and an attempt was made to enable a Data

 

 

 

AVailable event OR a DAV event is enabled and an attempt is made to set the

 

 

 

clock source to INTernal.

 

 

 

 

 

 

-222, “Data out of range”

Data for a parameter is outside of limits. For example, an attempt to set debounce

 

 

 

time to 10000 seconds.

 

 

 

 

 

 

-230, “Data corrupt or stale”

Data available (DAV) is FALSE. New data has not been clocked into the input

 

 

 

circuitry since the input circuitry has been reprogrammed or since the last current

 

 

 

value was read.

 

 

 

 

 

 

2025, “Invalid port number for access TYPE”

Port not 0 or 2 for LWORD access.

 

 

 

 

 

 

2026, ”Port number out of range”

Specified port is not 0, 1, 2, or 3.

 

 

 

 

 

 

2027, “Invalid bit number for access TYPE”

Bit number must be in the range of 0 to 15 for WORD and 0 to 31 for LWORD.

 

 

 

 

 

 

2029, “Event is not an OPERation status event”

The event specified is not DAVailable, NEDGe, or PEDGe.

 

 

 

 

 

 

2030, “Event is not an QUEStionable status event”

The event specified is not VDATa.

 

 

 

 

 

 

2031, “Bit number out of range”

Bit number must be between 0 and 15.

 

 

 

 

 

 

2032, “Totalizers not installed”

TOTalize bit set to 1 but totalizers are not installed on card.

 

 

 

 

 

 

2034, “Register is read only”

Attempt to write to a read only register.

 

 

 

 

 

 

3000, “Illegal while initiated”

Trigger system must be in idle state if the FIFO trigger is EVENt.

 

 

 

 

 

 

 

 

 

Error Messages 105