INPutn:CLOCk[:SOURce]?
Returns the programmed value of the input clock source for Port n.
Parameters
|
|
|
|
Parameter Name | Parameter Type | Range of Values | Default |
|
|
|
|
INPutn | numeric | 0, 1, 2, 3 | 0 |
|
|
|
|
|
|
|
|
Comments | • The value returned is an unquoted string of EXT or INT. | |
Example | INPut2:CLOCk[:SOURce]? | queries the input clock source for Port 2. |
INPutn:DEBounce:TIME <time> MINimum MAXimum DEFault
Programs the input debounce time for Port n. Ports 0 and 1 share the same debounce circuitry; Ports 2 and 3 share the same debounce circuitry.
Parameters
Parameter Name | Parameter Type | Range of Values | Default |
|
|
|
|
INPutn | numeric | 0, 1, 2, 3 | 0 |
<time> | numeric (floating pt) | 18.0 μSec through 9600 Sec |
|
DEFault |
| Default 18.0 μSec | 18.0μSec |
MINimum |
| Minimum 18.0 μSec |
|
MAXimum |
| Maximum 9600 Sec |
|
|
|
|
|
|
|
|
|
Comments • The available debounce times are:
18.0μS, 36 μS, 72 μS, 144 μS, 288 μS, 576 μS, 1.13 mS, 2.26 mS, 4.6 mS, 9.2 mS, 18.4 mS, 36.9 mS, 73.8 mS, 148.0 mS, 294.0 mS, 590.0 mS, 1.18 S, 2.36 S, 4.72 S, 9.43 S, 18.9 S, 37.8 S, 75.0 S, 150.0 S, 300 S, 600 S, 1200 S, 2400 S, 4800 S, 9600 S. Any value sent other than those listed is rounded to the nearest discrete value.
•Debounce time is rounded to the nearest discrete value. For example 16.0 μS to
18.4μS is rounded to 18 μS, 18.5 μS to 36.4 μS is rounded to 36 μS, 36.5 μS to 72.4 μS is rounded to 72 μS, etc.
•Ports 0 and 1 use the same debounce time, Ports 2 and 3 use the same debounce time. For n = 0 or n = 1, both Ports 0 and 1 are set; for n = 2 or n = 3, both Ports 2 and 3 are set.
•*RST Condition: sets the value for the debounce time to 18.0μSec for all four Ports.
HP E1459A SCPI Command Reference 47