SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

Upholstery has no rips.

Armrest pad sits flush against arm tube.

Inspect foam handgrips for damage. If damaged, have them replaced by a qualified technician.

Check that all labels are present and legible. Replace if necessary.

Inspect electrical components for signs of corrosion. Replace if corroded or damaged.

Check limit switch position for recline.

Make sure seating systems with the recline function have the retaining pin in place at the top of each back cane.

Make sure all systems, except Elevate Only, have the retaining pin in place at the back of each seat rail.

Check that spreader bar mounting fasteners are tight.

Check that back cane mounting fasteners and back mounting fasteners are tight.

Troubleshooting

NOTE: For additional troubleshooting information and explanation of error codes, refer to the wheelchair base owner’s manual and the individual Electronics Manual supplied with each wheelchair.

SYMPTOM

PROBABLE CAUSE

SOLUTIONS

 

 

 

Wheelchair Power On but does not

System tilted or reclined beyond

Return to neutral position (upright and

drive.

drive lock-out angle (20°).

completely lowered). Refer to

 

 

Operating Powered Seating

 

 

Systems on page 24.

 

 

Contact Invacare/Dealer for service if

 

Motor lock levers are

this does not solve the problem.

 

disengaged.

 

 

 

Engage motor lock levers. See

 

 

wheelchair base owner’s manual.

 

 

 

Seating system not functioning or

Low batteries.

Charge batteries.

working intermittently.

 

 

 

Faulty electrical connection.

Check all connections.

 

Seat has been driven under a

 

 

heavy load for an extended

Allow time for the electronics to cool

 

period of time.

down (Light Duty Use).

 

 

 

SPM L Park Brake Fault or

Motor lock levers disengaged

Engage motor lock levers.

(Error code E9 or E10).

 

SPM R Park Brake Fault displays

 

 

 

and wheelchair does not drive.

 

 

 

 

 

CHARGER PLUGGED IN displays.

Battery charger connected

Unplug battery charger from the

 

(Error code E28).

wheelchair.

 

 

 

Formula™ Powered Seating

20

Part No 1122170

Page 20
Image 20