SECTION
Adjusting Adjustable Angle Flip-Up Footplates
Depth Adjustment
NOTE: For this procedure, refer to FIGURE 6.9 on page 40.
1.Remove the two flat screws, washers and locknuts that secure footplate to the half clamp.
NOTE: Observe the angle of the footplate for reinstallation.
2. Move footplate to one of four mounting positions.
NOTE: If desired depth is still not obtained, rotate the half clamp on the footplate hinge 180°.
3. Retighten the two flat screws, washers and locknuts.
NOTE: The settings for positioning the footplates on the
Angle Adjustment
NOTE: For this procedure, refer to FIGURE 6.9 on page 40 and FIGURE 6.10.
1.Loosen, but DO NOT remove, the two flat screws, washer and locknuts that secure the footplate to the footplate hinge (FIGURE 6.9).
2.Position the footplate to the necessary angle to accommodate the user (FIGURE 6.10).
3.Retighten the two flat screws, washers and locknuts.
| Side View of | |
Footrest Support | Footplate and | |
Footplate | Footrest | |
Support | ||
|
FIGURE 6.10 Angle Adjustment
Perpendicular and/or Inversion/Eversion Adjustment
NOTE: For this procedure, refer to FIGURE 6.9 on page 40 and FIGURE 6.11.
NOTE: It is not necessary to remove the footplate to perform this adjustment.
1.Insert a flathead screwdriver through the half clamp on the footplate (FIGURE 6.9).
2.Slowly turn nylon adjustment screw in or out until footplate is perpendicular to the footrest assembly or the desired inversion or eversion is obtained (FIGURE 6.11).
Front View of Footplate
and Footrest Support
Footrest Support
Footplate
FIGURE 6.11 Perpendicular and/or
Inversion/Eversion Adjustment
Part No 1122170 | 41 | Formula™ Powered Seating |