SECTION 6—FRONT RIGGINGS

Adjusting Adjustable Angle Flip-Up Footplates

Depth Adjustment

NOTE: For this procedure, refer to FIGURE 6.9 on page 40.

1.Remove the two flat screws, washers and locknuts that secure footplate to the half clamp.

NOTE: Observe the angle of the footplate for reinstallation.

2. Move footplate to one of four mounting positions.

NOTE: If desired depth is still not obtained, rotate the half clamp on the footplate hinge 180°.

3. Retighten the two flat screws, washers and locknuts.

NOTE: The settings for positioning the footplates on the half-clamps may vary for each footplate.

Angle Adjustment

NOTE: For this procedure, refer to FIGURE 6.9 on page 40 and FIGURE 6.10.

1.Loosen, but DO NOT remove, the two flat screws, washer and locknuts that secure the footplate to the footplate hinge (FIGURE 6.9).

2.Position the footplate to the necessary angle to accommodate the user (FIGURE 6.10).

3.Retighten the two flat screws, washers and locknuts.

 

Side View of

Footrest Support

Footplate and

Footplate

Footrest

Support

 

FIGURE 6.10 Angle Adjustment

Perpendicular and/or Inversion/Eversion Adjustment

NOTE: For this procedure, refer to FIGURE 6.9 on page 40 and FIGURE 6.11.

NOTE: It is not necessary to remove the footplate to perform this adjustment.

1.Insert a flathead screwdriver through the half clamp on the footplate (FIGURE 6.9).

2.Slowly turn nylon adjustment screw in or out until footplate is perpendicular to the footrest assembly or the desired inversion or eversion is obtained (FIGURE 6.11).

Front View of Footplate

and Footrest Support

Footrest Support

Footplate

FIGURE 6.11 Perpendicular and/or

Inversion/Eversion Adjustment

Part No 1122170

41

Formula™ Powered Seating

Page 41
Image 41
Invacare Powered Seating manual Adjusting Adjustable Angle Flip-Up Footplates, Depth Adjustment, Angle Adjustment