SECTION
SYMPTOM | PROBABLE CAUSE | SOLUTIONS | |
|
|
| |
SPM Battery Fault displays and the | Batteries need to be charged | Charge batteries. If batteries fail to | |
wheelchair does not drive. | (Error code E14). | charge properly, check battery charger | |
| or replace batteries. Refer to Replacing | ||
|
| ||
|
| Batteries on page 63. | |
|
|
| |
JOYSTICK TIMEOUT displays and | Joystick or input device is | Turn off power, reconnect the joystick | |
the wheelchair does not drive. | disconnected (Error code 32). | of input device and turn power on. | |
|
| ||
|
|
| |
JS REV TOO LARGE | The joystick or input device is | Replace joystick or input device. | |
sending a value outside of the |
| ||
JS FWD TOO LARGE |
| ||
reverse, forward, left or right |
| ||
|
| ||
JS LFT TOO LARGE or | limits (Error codes E01, E02, |
| |
JS RGT TOO LARGE | E03 or E04). |
| |
|
| ||
displays and the wheelchair does not |
|
| |
drive. |
|
| |
|
|
| |
NEUTRAL TESTING displays. | The joystick neutral test has | Release the joystick and try to get the | |
| failed (Error code E18). | joystick back into the | |
|
| position. | |
|
|
| |
BAD JOYSTICK CAL VALUES | The joystick calibration values | Contact Dealer/Invacare for | |
are outside of the expected | recalibration of the joystick (joystick | ||
displays and the wheelchair does not | |||
range (Error code E19). | throw procedure). | ||
drive. | |||
|
| ||
|
|
| |
SPM NOT CONNECTED | The MPJ or Display module is | Check the connections between the | |
not communicating with the | joystick or display and the controller. | ||
| |||
| control module (Error code | Turn the power off and then back on. | |
| E200). | Replace the controller if necessary. | |
|
|
| |
SPM Communications Fault | The controller has determined a | Turn the wheelchair off and back on. | |
fault during a previous |
| ||
displays and the wheelchair drives |
| ||
process (Error code E41). |
| ||
slowly. |
| ||
|
| ||
|
|
| |
| The Proportional or Digital | This is normal behavior. | |
ATTENDANT ACTIVE and | Attendant control is active and |
| |
displays. | can be used to drive the chair |
| |
| (Error code W05). |
| |
|
|
| |
Batteries draw excessive current | Battery failure. | Have batteries checked for shorted cell. | |
when charging. |
| Replace if necessary. | |
| Electrical malfunction. | Contact Dealer/Invacare for service. | |
|
|
| |
Battery indicator flashes the charge | Battery failure. | Check batteries for shorted cell. | |
level is low - immediately after |
| Replace if necessary. | |
recharge. |
|
| |
| Malfunctioning battery charger. | Contact Dealer/Invacare for Service. | |
| Electrical malfunction. | Contact Dealer/Invacare for Service. | |
|
|
| |
Battery indicator flashes the charge | Batteries not charged. | Have charger checked. | |
level is low - too soon after being |
|
| |
recharged. | Weak batteries. | Replace batteries if necessary. Refer to | |
|
| Replacing Batteries on page 63. | |
|
|
| |
Motor “chatters” or runs irregular. | Electrical malfunction. | Contact Dealer/Invacare for Service. | |
|
|
|
Part No 1122170 | 21 | Formula™ Powered Seating |