25
6. Assembly
6.6 Light sources for the incident light axis
Caution!
Light sources pose a potential irradiation
risk (glare, UV-radiation, IR-radiation).
Therefore, lamps have to be operated in
closed housings.
Ensure that the lamp housing has been dis-
connected from the power supply. Unplug
the power plug and the power supply during
assembly.
During assembly work on xenon burners,
always wear the protective gloves and face
protection supplied (Fig. 15) (risk of explo-
sion).
Never touch the glass parts of the burner
with bare hands.
Never look directly into the beam path
(blinding hazard).
Lamp housing 106/106 z
This lamp housing is suitable for use with a 12 V
100W halogen lamp or a variety of gas discharge
lamps.
Caution!
Make sure to follow the instructions and
safety notes of the lamp supplier.
Before changing lamps allow it to cool down
for at least 30 min.!
Fig. 16 106/106 z lamp housing (on the side, open)
1Cover raised
2Collector
312 V 100 W lamp or
gas discharge lamp in mount
4Reflector (mirror)
5, 6, 7 Adjusting screw for x-y reflector
8Fastening screw for the lamp mount
9Socket for contact plug
1
2
3
898
4
5
6
7
Fig. 15
Protective gloves and mask