7. Startup

7.3The display (Leica DM4000 B/ DM4000 M/DM4500 P)

The screen displays the microscope’s current settings. The content of the display depends on the features of the individual microscope. In the first column, corresponding symbols indicate the type of information: contrast method, magni- fication, light intensity, diaphragms, light split- ting for photo tubes.

Please see the abbreviation index for a list of abbreviations used, p. 36f.

Contrast method

In the first row, you find an indication of the ac- tive light axis (incident light or transmitted light) of the current contrast method and, if used, the current filter cube.

The shutter status display for the transmitted light or incident light shutter:

￿

￿

Light intensity

 

 

The current brightness setting is graphically de- picted by a beam. In addition, the light intensity is specified in 20 (coarse) or 255 (fine) incre- ments p. 59.

￿

￿

Diaphragm

 

 

The values for the field diaphragm (FD) and the aperture diaphragm (AP) are indicated numeri- cally. The field diaphragm in the incident light may be either round or rectangular. Accordingly, the FD designation is set in parentheses or in brackets: (FD) or [FD].

Note:

When using a digital camera, angular field dia- phragms are recommended.

Transmitted light shutter is open

Transmitted light shutter closed

Incident light shutter is open (for fluorescence only)

Incident light shutter is closed (for fluorescence only)

+

Magnification

The current objective magnification (OBJ), sometimes followed by the re-magnification of the magnification changer*, appears along with the total magnification:

Σ= Objective x re-magnification x eyepiece

Light distribution

If a motorized tube is present, the light division between ocular (Eye) and photo port (Docu) is indicated in %.

Note:

The display may flash after the initialization phase or even during microscopy. This always occurs when the contrast method selected can not be performed with the microscopic settings. For example, an objective may be rotated in that is not suited to the contrast method chosen.

Then check your settings.

39

Page 39
Image 39
Leica DM5000B manual Startup Display Leica DM4000 B/ DM4000 M/DM4500 P, Contrast method, Magnification, Light distribution

DM4500P, DM5000B, DM4000M specifications

The Leica DM4000M and DM4000B are state-of-the-art microscopes designed for professional use in the fields of biology, materials science, and clinical applications. Renowned for their precision and innovative features, these instruments are perfect for researchers and clinicians needing high-resolution imaging capabilities.

One of the standout features of both models is the advanced motorized focusing system, which allows for swift adjustments and precise control. This feature is particularly useful in time-sensitive research environments, where accuracy and speed are paramount. The ergonomically designed focus mechanism promotes user comfort during prolonged observation sessions.

Both the DM4000M and DM4000B incorporate the revolutionary Leica Application Suite (LAS) software. This intuitive platform is designed to maximize the functionality of the microscope, enabling users to capture, analyze, and share images seamlessly. The software’s integrated tools are perfect for documenting findings and enhancing research productivity.

Another notable characteristic of the DM4000 series is the modular design, which allows for easy customization and upgrading. This aspect ensures that users can tailor their microscopes to meet specific research needs, whether it be for fluorescence microscopy, phase contrast, or even special imaging techniques like HSR or IR.

The high-performance optics provide exceptional image contrast and clarity, allowing users to observe minute details in samples. The combination of high numerical aperture objectives and advanced optical coatings enhances the resolution, making the DM4000 series ideal for examining intricate biological specimens as well as materials with complex textures.

The DM4000B model is particularly suited for routine clinical applications, featuring specific tools designed for rapid diagnosis and efficient workflows. Its user-friendly interface and dedicated clinical applications streamline laboratory processes, making it an essential device in any clinical setting.

Additionally, both models are equipped with LED illumination, which offers consistent light intensity and color temperature. This feature improves sample clarity while reducing heat generation, thereby protecting sensitive specimens during prolonged observation periods.

In conclusion, the Leica DM4000M and DM4000B microscopes represent a blend of advanced technology, intuitive design, and high adaptability. Their user-focused features and exceptional optical performance make them indispensable tools for researchers and clinical professionals aiming for excellence in microscopy.