Chapter4 SignalConnections
©NationalInstruments Corporation 4-45 6023E/6024E/6025EUser Manual
Figure4-36. GPCTR0_GATESignal Timing in Edge-Detection Mode
GPCTR0_OUT SignalThis signal is available only as an output on the GPCTR0_OUT pin. The
GPCTR0_OUTsignal reflects theterminal count (TC) of general-purpose
counter 0. You have two software-selectable output options—pulse on TC
andtoggle output polarity on TC. The output polarity is software-selectable
for both options. This output is set to high impedance at startup.
Figure 4-37 shows the timing of the GPCTR0_OUT signal.
Figure4-37. GPCTR0_OUTSignal Timing
GPCTR0_UP_DOWN SignalThissignal can be extern ally input on the DIO6 pin and is not available as
anoutput on the I/O connector. The general-purpose counter 0 counts down
whenthis pin i sat a logiclow and count up when it is at a logic high. You
can disable this input so that software can control the up-down
functionality and leave the DIO6 pin free for general use.
Rising-Edge
Polarity
Falling-Edge
Polarity
tw
tw=10 ns minimum
GPCTR0_SOURCE
GPCTR0_OUT
GPCTR0_OUT
(ToggleOutput on TC)
(Pulseon TC)
TC