AppendixA Specificationsfor PCI and PXI Buses
6023E/6024E/6025EUser Manual A-2 ni.com
Overvoltage protection
FIFObuffer size ......................................512S
Data transfers..........................................DMA, interrupts,
programmed I/O
DMA modes ...........................................Scatter-gather
(single transfer, demand transfer)
Configurationm emory size....................512 words
Accuracy Information
Signal Powered On Powered Off
ACH<0..15> ±42 ±35
AISENSE ±40 ±25
NominalRange (V)
AbsoluteAccuracy RelativeAccuracy
%of Reading
Offset
(mV)
Noise+ Quantization
(mV)
Temp
Drift
(%/°
°°
°C)
Absolute
Accuracy
atFull
Scale
(mV)
Resolution(mV)
Positive
FS
Negative
FS 24Hours 1Year SinglePt. Averaged SinglePt. Averaged
10 10 0.0872 0.0914 6.38 3.91 0.975 0.0010 16.504 5.89 1.28
55 0.0272 0.0314 3.20 1.95 0.488 0.0005 5.263 2.95 0.642
0.5 0.5 0.0872 0.0 914 0.340 0.195 0.049 0.0010 0.846 0.295 0.064
0.05 0.05 0.0872 0.0914 0.054 0.063 0.006 0.0010 0.106 0.073 0.008
Note:Accuracies are valid for measurementsfollowing an internal E Series calibration. Averaged numbers assumedithering and averaging of
100single-channel readings. Measurement accuracies arel istedfor operational temperatures within ±1 °C of internal calibration tem perature
and±10 °C of external or factory-calibration temperature. One-year calibration intervalrecommended. The Absolute Accuracy at Full Scale
calculationswere performed for a maximum range input voltage (for example, 10 V for the ±10 V range) after one year, assuming 100 pt