Cleaning and Low-Level Disinfection

The reusable sensors should be cleaned and disinfected, but never sterilized. Follow the procedure outlined below.

Warnings

Use only validated cleaning agents and disinfectants listed below; do not use any others. Failure to do so may damage the sensor or its con- necting wires and shorten product lifetime or cause safety hazards.

Select disinfectants carefully as some have very similar names but completely different compositions.

Do not immerse the sensor connector in any of the cleaning solutions, disinfectant, or other liquid (only the sensor and cable housing may be immersed, not the connector).

Do not soak sensors in disinfectants longer than specified by the dis- infectant manufacturer.

Do not sterilize the sensors.

Validated Cleaning Agents

- Mild Detergent - Salt Solution (1%)

Validated Disinfectants

Metricide® 28

Cidex® Formula 7

Kohrsolin® (2%)

Metricide® Plus 30

Cidex® OPA

Mucapur®-CD (1%)

Terralin® Liquid

Cidex® Plus

Isopropanol (70%) or

Incidin® Liquid

Omnicide® 28

Isopropanol Wipe (70%)

 

 

 

Step

Cleaning and Low-Level Disinfection

1Clean sensor following instructions supplied with cleaning agent.

2Disinfect sensor following instructions supplied with disinfectant.

3Rinse sensor in water, wiping it dry with a clean cloth, leaving to dry com- pletely. If you notice any deterioration or damage to sensor or cable, dis- pose the sensor immediately.