Technical Data

Warm up Period

15 minutes

Weight

 

Optical Head

100 g (3.5 oz)

Electronics Housing

710 g (9 oz)

3.3 Electrical Specifications

Power Supply

24 VDC ±20%, 500 mA (max 100 mV peak to peak of ripple)

Power Consumption

max. 12 W

Output Isolation

500 V AC or DC provided by manufacturer supplied power

 

supply accessory

Dielectric Withstand Voltage

500 V

Outputs

 

 

 

 

 

Analog

0 ‐ 20 mA, 4 ‐ 20 mA, 16 bit resolution

 

max current loop impedance: 500

Digital RS485

networkable to 32 sensors

 

Baud rate: 300, 1200, 2400, 9600, 19200, 38400 (default)

 

Adjustable baud rate only available through 2‐way RS485.

 

Data format: 8 bit, no parity, 1 stop bit,

 

Software selectable 4‐wire, full‐duplex non‐multidrop, point‐to‐

 

point or 2‐wire half duplex multidrop

Relay

Contacts max. 48 V, 300 mA, response time < 2 ms, (software

 

programmable)

Input

 

 

 

 

 

External Reset

TTL input, trigger for resetting peak or valley hold

 

 

 

 

 

 

 

 

 

 

Sensor

 

 

 

 

 

 

Trigger

GND

Figure 1: External Reset Wiring Diagram

6

Marathon Series FA/FR

Page 14
Image 14
RayTek FR, FA operating instructions Electrical Specifications