Maintenance

10 Maintenance

Our sales representatives and customer service are always at your disposal for questions regarding application assistance, calibration, repair, and solutions to specific problems. Please contact your local sales representative if you need assistance. In many cases, problems can be solved over the telephone. If you need to return equipment for servicing, calibration, or repair, please contact our Service Department before shipping. Phone numbers are listed at the beginning of this document.

10.1 Troubleshooting Minor Problems

Symptom

Probable Cause

Solution

 

 

 

No output

No power to instrument

Check the power supply

 

 

 

Erroneous temperature

Faulty sensor cable

Verify cable continuity

 

 

 

Erroneous temperature

Field of view obstruction

Remove the obstruction

 

 

 

Erroneous temperature

Window lens

Clean the lens

 

 

 

Erroneous temperature

Wrong slope or emissivity

Correct the setting

 

 

 

Temperature fluctuates

Wrong signal processing

Correct Peak/Valley Hold or Average settings

 

 

 

Table 8: Troubleshooting

Marathon Series FA/FR

45