Operation

6.4 Factory Defaults

To globally reset the unit to its factory default settings, press the c and d buttons at the same time for approximately 2 seconds. The baud rate will not change from the last value when this is done.

Parameter

FA (1-color unit)

FR (2-color unit)

Display mode

°C, TEMP- Display

2C, °C, TEMP- Display

Emissivity

1.00

 

1.00

 

Slope

-

 

1.000

 

PKH

0.0 s

 

0.0 s

 

AVG

0.0 s

 

0.0 s

 

VAL

0.0 s

 

-

 

Baud rate

38400

 

38400

 

Relay alarm output

controlled by unit

controlled by unit

Current Output

4 – 20 mA

 

4 – 20 mA

 

Temperature setting for 4 mA

FA1A:

475°C (887°F)

FR1A:

500°C (932°F)

 

FA1B:

800°C (1472°F)

FR1B:

700°C (1290°F)

 

FA1C:

1200°C (2192°F)

FR1C:

1000°C (1830°F)

 

FA1G:

750°C (1382°F)

 

 

 

FA2A:

250°C (482°F)

 

 

 

FA2B:

400°C (752°F)

 

 

Temperature setting for 20 mA

FA1A:

900°C (1652°F)

FR1A:

1100°C (2012°F)

 

FA1B:

1900°C (3452°F)

FR1B:

1500°C (2732°F)

 

FA1C:

3000°C (5432°F)

FR1C:

2500°C (4532°F)

 

FA1G:

1675°C (3047°F)

 

 

 

FA2A:

800°C (1472°F)

 

 

 

FA2B:

1700°C (3092°F)

 

 

Panel Control

unlocked

 

unlocked

 

Serial Communication

4-wire, non multidrop

4-wire, non multidrop

Transfer modus RS485

Burst-Modus

 

Burst-Modus

 

Output string

UTEI = temperature unit, 1C temperature,

UTEI = temperature unit, 2C temperature,

(RS485)

emissivity, internal temperature

emissivity, internal temperature

Figure 28: Factory Defaults

32

Marathon Series FA/FR

Page 40
Image 40
RayTek FR, FA operating instructions Factory Defaults