RayTek FR operating instructions Fail‐Safe Operation, Error Codes in 1‐Color Mode FA models

Models: FA FR

1 67
Download 67 pages 6.88 Kb
Page 54
Image 54

Maintenance

10.2 Fail‐Safe Operation

The Fail‐Safe system is designed to alert the operator and provide a safe output in case of any system failure. Basically, it is designed to shutdown the process in the event of a set‐up error, system error, or a failure in the sensor electronics.

The Fail‐Safe circuit should never be relied on exclusively to protect critical heating processes. Other safety devices should also be used to supplement this function!

When an error or failure does occur, the display indicates the possible failure area, and the output circuits automatically adjust to their lowest or highest preset level. The following table shows the values displayed on the LED display and transmitted over the 2‐way interface.

Symptom

Error Code

Priority

Heater control temperature over range

ECHH

1 (high)

Heater control temperature under range

ECUU

2

Internal temperature over range

EIHH

3

Internal temperature under range

EIUU

4

Temperature under range

EUUU

5

Temperature over range

EHHH

6 (low)

Table 9: Error Codes in 1‐Color Mode (FA models)

 

Condition

 

 

2-Color

 

 

1-Color

 

 

1-Color*

 

 

Priority

 

 

 

 

 

 

 

 

 

(wide band)**

 

 

(narrow band)**

 

 

 

 

 

 

Heater control temperature over range

 

 

ECHH

 

ECHH

 

ECHH

 

1 (high)

 

 

Heater control temperature under range

 

 

ECUU

 

ECUU

 

ECUU

2

 

 

 

Internal temperature over range

 

 

EIHH

 

EIHH

EIHH

3

 

 

 

Internal temperature under range

 

 

EIUU

 

EIUU

 

EIUU

4

 

 

 

Wide band detector failure

 

 

EHHH

 

EHHH

 

<temperature>

5

 

 

 

Narrow band detector failure

 

 

EHHH

 

<temperature>

 

EHHH

6

 

 

 

Energy too low

 

 

EUUU

 

<temperature>

 

<temperature>

7

 

 

 

Attenuation too high (>98%)

 

EAAA

 

<temperature>

 

<temperature>

8

 

 

 

Attenuation too high >95%

 

 

<temperature>

 

<temperature>

 

<temperature>

9

 

 

 

("dirty lens", relay will go to “alarm” state)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2-color temperature under range

 

 

EUUU

<temperature>

 

<temperature>

10

 

 

 

2-color temperature over range

 

 

EHHH

 

<temperature>

 

<temperature>

11

 

 

 

1-color temperature (wide) under range

 

 

<temperature>

 

EUUU

<temperature>

12

 

 

 

1-color temperature (wide) over range

 

 

<temperature>

 

EHHH

<temperature>

13

 

 

 

1-color temperature (narrow) under range

 

<temperature>

 

<temperature>

 

EUUU

14

 

 

 

1-color temperature (narrow) over range

 

<temperature>

 

<temperature>

 

EHHH

 

15 (low)

*only available through RS485

**Wide and narrow band stands for the first and the second wavelength in 2-color mode

Table 10: Error Codes in 2‐Color Mode (FR models)