Sensor Location

4.5 Sensor Placement (1‐Color Mode)

Optical head placement for one‐color temperature measurements is more critical than two‐color measurements. The sensor must have a clear view of the target. There can be no obstructions on the lens, window, or in the atmosphere. The distance from the target can be anywhere beyond the minimum requirements, as long as the target completely fills the field of view. The following figure illustrates proper placement when using the one‐color mode.

best good incorrect

Target greater than spot size

Target equal to spot size

Target smaller than spot size

Figure 10: Proper Sensor Placement in 1‐Color Mode

4.6 Sensor Placement (2‐Color Mode)

The following figure shows head placement under various conditions where two‐color temperature measurements can be taken. Note, however, that if the sensor signal is reduced more than 95% (including emissivity and obscuration of the target), the sensor accuracy also degrades.

14

Marathon Series FA/FR

Page 22
Image 22
RayTek FR, FA operating instructions Sensor Placement 1‐Color Mode, Sensor Placement 2‐Color Mode