Maintenance

Turn 1.3 mm (0.050“) hex wrench counter clockwise until cable is loose

Put cable out

Figure 36: Removing the Fiber optic Cable from the Optical Head

10.4.1.2 Removing the Fiber Optic Cable from the Electronics Housing

Complete the following steps to disconnect the fiber optic cable from the electronics housing:

1.First loosen the cable connecting sleeve.

2.Loosen the cable receptacle screw to release the cable.

3.Pull cable from electronics enclosure, and immediately place a protective cap over the end of the fiber optic cable. Do not use adhesive tape on the cable end.

First loosen cable connecting sleeve

Then loosen screw and pull cable from coupling.

Figure 37: Removing the Fiber optic Cable from the Electronics Housing