RayTek FA, FR Current Output Values in accordance to an Error, Examples of failsafe conditions

Models: FA FR

1 67
Download 67 pages 6.88 Kb
Page 55
Image 55
EUUU
C T0999
2 to 3 mA abnormal state

Maintenance

The relay is controlled by the temperature selected on the display. If any failsafe code appears on the display, the relay changes to the “abnormal” state. The exception is the “dirty window” condition. This causes the relay to change state, leaving a normal numerical temperature output. The dirty window is detected in either 1‐color or 2‐color mode.

 

Error Code

0 – 20 mA Output

 

4 – 20 mA Output

 

 

no error

according to temperature

 

according to temperature

 

 

ECHH

21 to 24 mA

21 to 24 mA

 

 

ECUU

0 mA

2 to 3 mA

 

 

EIHH

21 to 24 mA

21 to 24 mA

 

 

EIUU

0 mA

2 to 3 mA

 

 

EUUU

0 mA

2 to 3 mA

 

 

EHHH

21 to 24 mA

21 to 24 mA

 

 

EAAA

0 mA

 

2 to 3 mA

 

Table 11: Current Output Values in accordance to an Error

If two errors occur simultaneously, the higher priority error is the one that is presented on the LED’s digital and analog outputs. For example, in 2‐color mode, if the internal ambient is too high and the attenuation is too high, the unit outputs EIHH on the LED’s and digital output and 21 mA on the analog output. However, since 2‐color wide band and narrow band temperatures may all be presented simultaneously through RS485, their over and under range conditions are independent.

Examples of failsafe conditions:

1.One‐color temperature is selected for display on the LED’s. Two‐color temperature is transmitted in burst mode. Wide band temperature is under range. Two‐color temperature is 999°C.

Outputs:

Display:

RS485:

Analog:

Relay:

2. Two‐color temperature is selected for display on LED’s. All three temperatures are transmitted in burst mode. Two‐color temperature is 1021°C. Wide band temperature is 703°C. Narrow band temperature is 685°C. Attenuation is above 95%, the “dirty window”

threshold.

 

Outputs:

 

Display:

1021

RS485:

C T1021 W0703 N0685

Analog:

scaled to temperature, between 4 and 20 mA

Relay:

abnormal state

Marathon Series FA/FR

47

Page 55
Image 55
RayTek FA, FR operating instructions Current Output Values in accordance to an Error, Examples of failsafe conditions